Each instrument will have a unique set of commands. It sounds to me that you should be using IVI drivers.
You are misunderstanding SCPI. The implementation of the exact commands is vendor dependent. They can vary quite a bit from vendor to vendor though from instrument to instrument from the same vendor, Leeds so. The manual must be referenced.